Centrifuge Operation
In this section, we will examine centrifuge operations.
A safety note, make sure to plug in the centrifuge before turning on the power pack.
![](/web/image/302257-323efffe/36084.webp)
![](/web/image/302258-04ffea3c/36085.webp)
To operate the centrifuge first remove the rotor cover, when taking Accu-tubes in or out, the rotor cover should be placed inside the centrifuge lid. This will remind you to always replace the rotor cover before spinning samples.
![](/web/image/302259-e1decc5a/36086.webp)
![](/web/image/302260-b2a97ed2/background_36087.webp)
![](/web/image/302261-892a2e93/36088.webp)
Once a sample is loaded into the Accu-tube place the tube in one of the numbered slots. The rotor is numbered to assist with balancing the tubes.
![](/web/image/302262-ea63e7f9/36089.webp)
![](/web/image/302263-e04b6bfa/36090.webp)
Always remember to place tubes opposite each other on the rotor.
![](/web/image/302264-c060beaa/36091.webp)
If there is an odd number of tubes an empty Accu-tube must be used to balance the centrifuge.
![](/web/image/302265-37b93016/36092.webp)
Once the Accu-tubes are loaded, replace the rotor cover, the rotor cover must be tightened carefully. If it is not secure the Accu-tubes could break, if tightened too much the rotor cover could strip. Use approximately six turns for an appropriate fit.
![](/web/image/302266-0b125c8b/36093.webp)
Close the lid of the centrifuge.
![](/web/image/302267-73e5534e/36094.webp)
The on/off button is used to start the spin, this button is also used to abort a spin if for instance the rotor cover is not secured properly.
![](/web/image/302268-f27d3afc/36095.webp)
The centrifuge will operate at 12,000 revolutions per minute or RPMs for 5 minutes.
![](/web/image/302271-3c67ab24/36096.webp)
Regulations require that the RPM’s and timer be checked every 6 months. Checking the timer is as simple as running the centrifuge and comparing it with an external timer for the full 5 minutes.
![](/web/image/302269-06e506c5/36097.webp)
![](/web/image/302270-53e3b0bf/36098.webp)
When the spin is complete the lid will automatically open. The Accu-tube should be removed immediately and placed upright in the workstation until they are to be tested in the auto-reader. Storing the Accu-tubes in this upright position prevents the cell layers from mixing together.
For accurate results, tests must be run within 4 hours of the centrifuge spin.
Watch the full video version below:
Check out these links below to shop for the Drucker Diagnostics QBC AutoRead Plus and its accessories on Block Scientific Store: