Beckman DxC 700 AU Chemistry Analyzer

https://blockscientific.odoo.com/web/image/product.template/83474/image_1920?unique=fb6dd2c

Brand-new, used, and refurbished Beckman DxC units available.

User-friendly software, reliable, high throughput analyzer for mid to high-volume clinical labs.

Block Scientific provides expert installation and repair services.

Ask us about our multi-year service contracts, reagent contracts, and selling your old instruments.

0.00 0.0 USD 0.00

0.00

    This combination does not exist.

     Easily reorder reagents and American-made parts.
     Enjoy the fair price promise.
     Take pressure off your budget (and yourself)

                  

    Product Code: B98654

    Manufacturer: Beckman Coulter

    Need replacement parts for your DxC 700 AU? Click here to find them quickly.

    Don’t miss your DxC 700 AU opportunity!

    We currently have two Beckman DxC 700 AU analyzers available. Both were in active use until a few weeks ago.

    Years of Service: since 2020

    Condition: Maintained, recently active, in working condition

    Availability

    • 1 DxC 700 AU complete with hard drive
    • 1 DxC 700 AU without hard drive, cycle count of 886,499

    If you’ve planned to purchase a Beckman DxC in the future, this is your chance to stretch out that budget and upgrade other areas of your lab at the same time.

    Check out the images of the actual analyzer to the right.

    They are packaged and ready for immediate dispatch.

    Request a quote now


    DxC 700 AU with hard drive. In service since 2020, well maintained and in working condition.

    DxC 700 AU analyzer. In service since 2020, well maintained and in working condition. Hard drive has been removed to protect patient data.

    Low Cost of Ownership DxC 700 AU for Mid to High-Volume Labs

    The software that drove you nuts on old workhorse Beckman chemistry analyzers is gone. 

    In its place, on the DxC 700 AU, you’ll find simplicity. No headaches. 

    Choose this chemistry analyzer today for:


    Not the traditional Beckman AU series

    There’s a cooled stat compartment.

    You can load reagents while the DxC 700 AU is running.

    There’s a 150 sample continuous rack loader. 

    And did we mention the famously challenging AU-series software is gone?

    Your operators will love the small learning curve, and your department will recover the budget quickly via increased capacity.

    Get your quote within 90 minutes today


    DxC 700 AU troubleshooting is easy

    Waste less budget on external service

    Most functions are customer defined, so engineer intervention
    is rarely needed


    The DxC 700 AU has the highest uptime and longest average
    repair intervals in its cost class. 


    If you’re validating against a Siemens Vista 500, the most common comparable
    analyzer in the class, your DxC 700 AU cost will be lower and results comparable.


    Read the independent study here: Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer.

     Hot Beckman DxC 700 AU tip

    Give your lab the best potential uptime by pairing your system with a high-quality DI water system. The DxC 700 is thirsty, but performs better than most in its class when you keep it hydrated.

    Fewer errors with DxC 700 AU

    It’s not the heat, it’s the humidity!

    Is your environment humid or dry?

    The DxC 700 AU is best suited for humid environments, where it routinely outperforms dry chemistry analyzers in its class (errors with the DxC 700 AU vs dry chemistry analyzers in humid environments are significantly fewer).

    Click here and your quote will arrive within 90 minutes

     
     


    DxC 700 AU lowers your cost per test

    By investing in a Beckman DxC 700 AU, you also reduce the cumulative cost of small items, like disposable cuvettes.

    Reduce the time your operators spend on inventory tasks.

    Reduce the time wasted on replacing parts like:

    • Sample and reagent probes
    • Mixers
    • Syringes
    • ISE electrodes 
    • Optical lamp

    (Every replacement is done with 3 or fewer steps, and in less than 60 seconds).

    Then, switching from your current reagent supplier to Block Scientific saves you 10-15% on DxC 700 AU reagents instantly. 

    Don’t believe us? Ask for a quote here and you’ll have the proof within 90 minutes.

    Other suppliers use tiered pricing to give the biggest brands the biggest discounts because of their purchasing volume. We don’t do that. Instead, every customer and labs of every size get the same deals. It’s pricing 9 out of 10 labs couldn’t qualify for anywhere else.

    See your cost in 90 minutes with a quote


    Use your old instruments to fund your upgrade

    Sell us your old instruments and the cost of your Beckman DxC 700 AU becomes even more manageable.

    Here’s what you get with us:


    We give you a competitive value on your equipment based on set parameters.


    You get a clear payment timeframe and no hidden clauses.


    We assist with the decontamination plus we arrange and pay for all shipping charges. Your process is hassle free.

    Finance your investment to manage your lab’s budget heroically

    Financing options are available for every DxC 700 AU purchase, so small and medium sized laboratories can manage the burden of capital equipment purchases.

    Contact a member of our team to learn more

    DxC 700 AU service & repairs

    Nationwide, onsite service is available (even though your DxC 700 AU will rarely need it). We can give you:

    • Customized on-site annual service contracts.
    • Professional installation.
    • Equipment familiarization following your installation.
    • Time and material service visits, if your DcX 700 AU is not under our service contract.



    We want you to be happy and less stressed at work

    Block Scientific has served clinical labs across the United States and around the world for more than 40 years. We know how hard your days can be. So let us help you make them easier.

    Your Beckman DxC 700 AU is ready to ship when you need it. 

    We also offer a 30-day satisfaction guarantee on every chemistry analyzer we carry. So try it. There’s no risk.

    We’ll send you a quote within 90 minutes today


    Instruments Specifications

    • Newly designed intuitive graphical user interface, including simplified reagent inventory, calibration and quality control procedures
    • Online maintenance documentation
    • User-customized menu and free-text notepad
    • Colour alerts to highlight system operating conditions
    • Reliability for greater uptime with quick and easy replacement of common parts
    • No special tools required
    • No more than three steps and no longer than 60 seconds for replacement of parts, such as sample and reagent probes, mixers, syringes, ISE electrodes and optical lamp
    • Extensive online help with instructional videos
    • Available integration with pre-analytical automation through direct track sampling from Power Express
    • Available connection to Command Central with REMISOL Advance, allowing access to multiple instruments and automation consoles from a single workstation, optimizing laboratory management and improving decision making
    • PROService remote diagnostic capabilities ensure 24/7 continuous monitoring of vital system functions
    • Cooled STAT compartment, providing one-button STAT interrupt, advanced auto QC and calibration capabilities
    • Ability to load reagents while the analyzer is in measure mode
    • Priority sample repeat/reflex
    • Economical ISEs with long onboard stability; easy to maintain (only individual electrode replacement required)
    • 150-sample continuous rack loader


    More Specifications



    Analytical system
    Fully automated, random-access clinical chemistry system with STAT capability

    Analytical principles
    Spectrophotometry and potentiometry

    Assay types
    Endpoint, rate, fixed point and indirect ISE

    Analytical methods
    Colorimetry, turbidimetry, latex agglutination and indirect ISE

    Test menu applications
    I20

    Programmable tests
    117 photometrics (113 + LIH + HbA1c (Thb, HbA1c + HbA1c%)) and ISE

    Onboard parameters
    60 photometric tests and 3 ISEs (Na, K, Cl)

    Throughput
    800 photometric tests per hour, up to 1,200 with ISE
    ISE sample throughput: 200 per hour
    ISE maximum tests per hour: 600 if ISE only

    Sample types
    Serum, plasma, urine, whole blood (HbA1c) and other fluids

    Sampler capacity
    Rack sampler: 10 samples per rack (barcodes on primary tubes and on racks), capacity for 150 samples
    Refrigerated STAT carousel: 22 samples can be run simultaneously (cal, QC and routine samples)

    Sample tubes
    Primary and secondary tubes: diameter between 11.5 mm and 16 mm; height between 55 mm and 102 mm; nested micro cups

    Sample volume
    μL-25 μL in 0.1 μL increments

    Sample quality analysis
    Lipemia, hemolysis and icterus indices
    Clot detection and probe crash protection

     

    Sample bar-code formats
    NW7, EAN13, CODE 39, CODE 128, ISBT-128, 2 of 5 standard, 2 of 5 interleaved mixed readable (max. four types at the same time, except if using ISBT-128)

    Reagent supply
    60 positions for R1, 48 positions for R2 (refrigerated 4°C-12°C); bottle sizes: 15 mL, 30 mL, 60 mL

    Reagent volume
    R1: 10 μL-250 μL; R2: 10 μL-250 μL (1 μL increments)

    Total reaction volume
    120-350 μL

    Reaction cuvette
    Permanent glass cuvettes

    Reaction time
    Up to 8 minutes, 33 seconds

    Reaction temperature
    37°C ±0.3°C

    Reaction method
    Dry bath

    Photometric range
    0 OD-3.0 OD

    Wavelength
    13 different wavelengths between 340 nm and 800 nm

    Calibration
    Auto-calibration, advanced calibration and cooled calibrator positions
    Calibrator value assignment by 2D barcode; 200 calibrators can be programed
    History of graphical calibration data stored

    Quality control
    Westgard Rules, Twin Plot and Levey-Jennings graphs, auto QC, cooled QC positions
    100 controls can be programed, 10 levels per test

     

    Reflex testing
    User-defined
    Automated sample pre-dilution
    Repeat run with increased or decreased sample volume or sample pre-dilution (3, 5, 10, 15, 20, 25, 50, 75, up to 100 times)

    Online
    Uni- and bi-directional host query communications

    Operating system
    Windows 7

    Data storage
    Up to 100,000 patient samples; reaction monitor: 400,000 tests, 300 index

    INSTALLATION REQUIREMENTS
    Dimensions (W x H x D)

    Analyzer:
    49 in x 51 in x 35 in (1250 mm x 1300 mm x 890 mm) 1,015 lbs (460 kg)

    Sampler:
    29 in x 37 in x 41 in (730 mm x 940 mm x 1040 mm)
    331 lbs (150 kg)

    Power supply
    200V, 208V, 220V, 230V, 240V, 50 Hz, 60 Hz, 3.8 kVA

    Water supply information
    Mean water consumption: 28 L per hour
    Water type: deionized CAP Type II, bacteria free
    Continuous flow supply
    Resistivity: less than 2.0 μS/cm filtered with a 0.5 μm filter

    Temperature and humidity
    18°C to 32°C, 20% to 80% RH (no condensation)

    Drain requirements
    Built-in waste pump
    Drain required: maximum height from floor < 59 in (< 1.5 m)