Dynex Agility Automated ELISA Systems
High throughput, fully automated Dynex Agility ELISA machine. Made in the USA.
Tell us your monthly, quarterly, or annual test volume and we'll include a quote on reagents too, so your reagent budget is predictable and easy to manage.
Get the total cost of ownership for your Dynex Agility in 90 minutes.
Product Code: DA-AES-1
2.5x Faster throughput than the conventional or manual method of many assays
Looking to upgrade from an old system or manual process to an automated ELISA testing system?
Dynex Agility ELISA processing systems are ideal cost-conscious options for labs that need to reduce human error and increase throughput.
1. Maintenance is easy
2. Operators prefer automated to manual assays
3. labs can increase throughput, reduce human error, and increase profitability by switching from manual to automated Dynex Agility ELISA machines.
Tell us your monthly, quarterly, or annual test volume and we'll include a quote on reagents too, so your reagent budget is predictable and easy to manage.
Why upgrade from manual assays to the fully automated Dynex Agility?
Upgrading to an automated ELISA system offers numerous advantages that make every day in the lab less stressful and more productive.
![](/web/image/301077-5177333d/tool%20copy.jpg?access_token=6b6c2552-9c15-4e4b-ba89-61c3252bf9d3)
Reduced hands-on time
This allows technicians to focus
![](/web/image/301083-31c19b7e/bar-chart%20copy.webp?access_token=a41c8d63-059c-4afd-bf2f-f5f334de12ec)
Increased throughput
Enable the processing of a higher number of samples simultaneously via automation, increasing overall lab productivity and enabling more tests to be run in less time.
![](/web/image/301086-d7b1022e/refresh-ccw%20copy.webp?access_token=3e4801ba-e8fd-4200-8fff-04a174add974)
Consistency
Ensure precise pipetting, consistent incubation times, and uniform washing, reducing variability and enhancing the reliability of results.
![](/web/image/301084-18c9e085/x-circle%20copy.webp?access_token=72c17ba6-fd28-4de1-a054-aca29b14dbe9)
Minimize errors
Minimize human errors, such as pipetting inaccuracies and timing mistakes, leading to more consistent and accurate data.
![](/web/image/301087-e1e862fa/alert-triangle%20copy.webp?access_token=93bbfefb-181c-406a-8bb1-ac023b1cf7c0)
Real-time alerts
Real-time alerts for gross-pipetting errors and automated result calculations ensure high-quality outputs and reduce the need for repeat tests.
![](/web/image/301085-6ecfc7c5/git-pull-request%20copy.webp?access_token=5e3d9379-7f26-4260-8406-741b3aa099e7)
Traceability and verification
Built-in traceability and verification processes ensure compliance with regulatory standards and provide peace of mind.
Request a quote today to see how the Dynex Agility can make your lab easier to work in and easier to run.
Request a quote today to see how the Dynex Agility can make your lab easier to work in and easier to run.
We use one of these Dynex ELISA machines and I can say they are very helpful especially because we have more than 70 probes a day.
One analysis takes between 1-4 hours depending on the type of analysis but you only have to mount the reagents in the beginning and everything else is automatic from pipetting to measuring to analyzing.
I almost never run into any problems with the system which cannot be solved immediately and maintaining the system is not really time-consuming nor difficult.
The probe preparation is not difficult either.
I would say the system is user-friendly and helpful. Compared to our legionella assay, which we do manually, an automated system is extremely helpful.
- Med Lab Technician
A Dynex Agility ELISA machine lowers your operational costs
Automated ELISA processes reduce human error, improve result consistency, and increase throughput.
Your lab can also save significant consumable costs and optimize the use of your human resources with the Dynex Agility system.
Save on reagents
Optimize the use of reagents and consumables, so your lab saves on monthly expenses.
Reduce labor costs
Less manual intervention means your technicians are less likely to get operator burnouts. Instead, they can focus their energies on activities that require human interpretation.
Future-proof your operations
Handle various assays and volumes, offering flexibility and scalability to meet changing lab demands.
Customize your solution
Handle various assays and volumes, offering flexibility and scalability to meet changing lab demands.
Get your quote and ask us to include reagents (tell us your monthly, quarterly, or annual volume) to make the annual cost of ownership of your Dynex Agility predictable.
Request a quote on your Dynex Agility now
Dynex Agility and other Dynex ELISA analyzer models
Dynex Agility - for busy labs that need to reduce manual assays
Designed for high-volume processing, the Dynex Agility system lets you run multiple simultaneous ELISA tests with long walk-away time.
It reduces hands-on time by up to 60% and increases your current ELISA testing capacity.
If your lab needs to scale up throughput, you can recover the cost of your Dynex Agility within 2 years via your increased capacity.
![](/web/image/301090-a956300a/Dynex%20Ds2%20_no%20bg.webp?access_token=c686be68-f136-438d-a78a-f894c6c8c603)
Dynex DSX - multi-tasking analyzer with medium throughput
The DSX system handles various assays and volumes simultaneously, delivering the speed and accuracy your lab demands.
Advanced automation features at every stage of the workflow provide the walk-away productivity that modern laboratories need.
![](/web/image/301091-1d765152/Dynex%20DXS.webp?access_token=6d7c22b1-c638-4eed-9a8f-ba0a734d3fe7)
Dynex DS 2 - Compact, small footprint automation ELISA processor
The DS2 system is perfect for smaller labs with limited space.
Despite its small footprint, the DS2 significantly enhances productivity and ensures fast and accurate results.
![](/web/image/301093-f7cedcc6/Dynex%20Agility.webp?access_token=9a0f28bb-3812-4e7a-8fac-da866cb8fbbc)
Turn your old instruments into new budget!
Sell us your instruments and the cost of your Dynex Agility becomes even more manageable.
We take all the burden out of selling your unneeded lab equipment.
![](/web/image/300983-95325e33/guarantee-%281%29.png)
We give you a competitive value on your equipment based on set parameters
![](/web/image/301839-ea3871fa/safe-%281%29%20%281%29.webp?access_token=880013ac-bb86-42fb-83da-77709a97c691)
You get a clear payment timeframe and no hidden clauses
![](/web/image/301838-cda3f685/hasslefree-%281%29%20%281%29.webp?access_token=c41bfc4a-2006-4e7f-96f5-85c851e1490d)
We assist with the decontamination plus we arrange
and pay for all shipping changes. Your process is hassle free.
Dynex Agility
Physical Specifications | ||
---|---|---|
Dimensions | Metric | Non Metric |
Width: | 125 cm | 50 in |
Depth: | 90 cm | 36 in |
Height: | 123 cm | 49 in |
Footprint: | 1.19 sq m | 12.5 sq ft |
Bench Weight: | 213 kg (max.) | 469 lbs (max.) |
Ship Weight: | 296 kg (max.) | 650 lbs (max.) |
Noise Output: | < 70dB |
Power Supply Requirements | |
---|---|
Voltage | 100 – 240 V automatic conversion |
Frequency | 50/60 Hz |
Power Consumption | Typically 400 VA (online UPS required) |
System Specifications | |
---|---|
Number of Plates | 12 (additional continuous load capability |
Sample Capacity | 200 (additional continuous load capability) |
Continuous Sample Loading | Yes |
Sample Tube Size | 10 mm - 17 mm diameter / 45 mm - 100 mm height |
Number of Reagents | Up to 16 different SmartKit® reagent packs simultaneously (additional packs can be added using continuous load capability) |
Number of Standard/Control Bottles | Included in each SmarKit reagent pack |
Dilution Wells Capacity | Up to 384 deep-well strips (48 x 8) |
Sample Tip Capacity | Up to 2,240 (20 × 112 with additional continuous load capability) |
Reagent Tip Capacity | Up to 98 tips (additional continuous load capability) |
Reader Specifications | |
---|---|
Photometric Range | 0.000 - 3.500 OD |
Spectral Range | 405 nm - 690 nm |
Filter Slots | 6 wavelength |
Precision | ±0.010 0D at 0.000 - 0.500 OD < < 1% CV at 0.501 - 2.000 OD < < 1.5% CV at 2.001 - 2.500 OD |
Accuracy | ±0.01 or 2.5% (0.000 - 3.000 OD) whichever is greater |
Read Time | < 30 seconds, single wavelength < < 50 seconds, dual wavelength |
Washer Specifications | |
---|---|
Manifold Configuration | 8-way wash head |
Programmable Volumes | 50 μL – 999 μL |
Wash Buffer Capacity | 4 wash bottles at 3.0 L with quantitative level-sensing |
Residual Wash Volume | < 3 µl per well with dual-axis sweep in a flat-bottom well |
Dispense Precision | ≤ 5% CV (with 300 µl in a 96 well plate) |
Waste Container | 10 L with quantitative level-sensing |
Incubator Specifications | |
---|---|
Number of Incubators: | 12 (6 ambient/6 elevated) |
Temperature Range: | RT + 4° C to 45° C (elevated incubator) < RT + 4° C (ambient incubator) |
Temparature Accuracy | ± 1°C |
Shaking | 14 Hz periodic or continuous |
Temperature Monitoring | Yes |
Regent Pipetting Specifications | |
---|---|
Reagent Tip Size: | 1.200 µL |
Reagent Pipetting Volume | 20 µL - 1,200 µL |
Reagent Pipetting Precision: | ≤ 3% CV at 10 shots at any volume in operating range |
Reagent Pipetting Accuracy: | ±2% of target volume (single-shot mode) |
Sample Pipetting Specifications | |
---|---|
Sample Tip Size | 300 µL |
Sample Pipetting Volume: | 10 µL - 300 µL |
Sample Pipetting Precision | ≤ 3% CV (at any operating volume above 10 µL) |
Time to Dispense | 11 minutes (typical) 50 µl of 96 samples to plate from sample tubes or deep-well plates |
Sample Pipetting Accuracy | ±2% of target volume |
Dilution Range | 1 part in 190 one-stage dilution < 1 part in 36,100 two-stage dilution |
Process Security | |
---|---|
Liquid Level Sensing | Yes (reagents, controls and samples) |
Level Sensor System | Pressure Differential |
Clot Detection | Yes |
Dispense Anomaly Detection | Yes |
Tip Detection | Yes |
Electronic Signature Pipetting | Yes |
Alarms | Yes |
Software | |
---|---|
Controlling Software | Agility Software |
Work Protocols (Assays): | Unlimited |
Data Processing | Quantitative & qualitative |
Levey-Jennings | Yes |
Westgards Rules | Yes |
Process Reporting | Event log + Error log |
Automatic Error Recovery | Yes |
Password Access Control | Yes |
Attributes and specifications subject to change without notice.
Dynex DSX
Physical Specifications | ||
---|---|---|
Dimensions | Metric | Non Metric |
Width: | 106 cm | 42 in |
Depth: | 91 cm | 36 in |
Height: | 80 cm | 32 in |
Footprint: | 0.98 sq m | 10.5 sq ft |
Bench Weight: | 136 kg (max.) | 300 lbs (max.) |
Ship Weight: | 244 kg (max.) | 537 lbs (max.) |
Power Supply Requirements | |
---|---|
Voltage | 100 – 240 V automatic conversion |
Frequency | 50/60 Hz |
Power Consumption | < 800 VA (online UPS recommended) |
System Specifications | |
---|---|
Number of Plates | 4 |
Sample Capacity | 96 (98 positions available) |
Continuous Sample Loading | Yes |
Sample Tube Size | 10 mm – 16 mm diameter / 40 mm – 100 mm height |
Number of Reagents | Up to 24 (25 mL tubes) |
Number of Standard/Control Bottles | 33 (1.5 mL vials) |
Assays per Plate | 1 assay per strip or up to 12 assays per plate |
Dilution Wells Capacity | Up to 192 deep-well strips (24 x 8) |
Sample Tip Capacity | 432 (4 x 108) |
Reagent Tip Capacity | 41 tips |
Reader Specifications | |
---|---|
Photometric Range | 0.000 – 3.000 OD |
Spectral Range | 405 nm – 690 nm |
Filter Slots | 6 wavelength |
Precision | ± 0.010 OD at 0.000 – 0.500 OD < 1% CV at 0.501 – 2.000 OD < 1.5% at 2.001 – 2.500 OD |
Accuracy | ± 0.01 OD or 2.5% (0.000 – 3.000 OD) whichever is greater |
Read Time | < 10 seconds, single wavelength < 20 seconds, dual wavelength |
Washer Specifications | |
---|---|
Manifold Configuration | 8-way |
Programmable Volumes | 50 μL – 999 μL |
Wash Buffer Capacity | 4 wash bottles at 2.0 L, with level-sensing |
Residual Wash Volume | < 3 μL per well with dual-axis sweep in a flat-bottom plate |
Dispense Precision | ≤ 5% CV (with 300 μL in a 96 well plate) |
Waste Container | 8 L with waste full sensor |
Incubator Specifications | |
---|---|
Number of Incubators: | Up to 4 |
Temperature Range: | RT + 7°C to 50°C (elevated incubator) <RT + 5°C (ambient incubator) |
Temparature Accuracy | ± 1°C |
Shaking | > 15 Hz periodic or continuous |
Temperature Monitoring | Yes |
Regent Pipetting Specifications | |
---|---|
Reagent Tip Size: | 1,300 µL |
Reagent Pipetting Volume | 25 μL – 1,000 μL |
Reagent Pipetting Precision: | ≤3% CV at 10 shots at any volume in operating range above 50 µL |
Reagent Pipetting Accuracy: | ± 2% of target volume at 50 µL or greater in operating range (single-shot mode) |
Sample Pipetting Specifications | |
---|---|
Sample Tip Size | 300 µL |
Sample Pipetting Volume: | 10 µL – 250 µL single-shot / 25 µL – 100 µL multi-shot |
Sample Pipetting Precision | ≤ 3% CV at any operating volume above 10 µL (single-shot mode) |
Estimated Cycle Time for Sample Pickup to Delivery on Plate | < 8 seconds |
Time to Dispense | 19 minutes (typical) < 50 µL of 96 samples to plate from sample tubes or deep-well plates |
Sample Pipetting Accuracy | ± 2% of target volume at any operating |
Sampling Time w/ Dilutions | < 26 minutes (typical) |
Dilution Range | 1 part in 190 one-stage dilution < 1 part in 36,100 two-stage dilution |
Process Security | |
---|---|
Liquid Level Sensing | Yes (reagents, controls and samples) |
Level Sensor System | Pressure Differential |
Clot Detection | Yes |
Dispense Anomaly Detection | Yes |
Tip Detection | Yes |
Electronic Signature Pipetting | Yes |
Alarms | Yes |
Software | |
---|---|
Controlling Software | REVELATION DSX® |
Work Protocols (Assays): | Unlimited |
Data Processing | Quantitative & qualitative |
Levey-Jennings | Yes |
Westgards Rules | Yes |
Process Reporting | Event log + Error log |
Automatic Error Recovery | Yes |
Password Access Control | Yes |
Attributes and specifications subject to change without notice.
Dynex DX2
Physical Specifications | ||
---|---|---|
Dimensions | Metric | Non Metric |
Width: | 54 cm | 21 in |
Depth: | 68 cm | 27 in |
Height: | 66 cm | 26 in |
Footprint: | 0.36 sq m | 3.9 sq ft |
Bench Weight: | 48 kg (max.) | 105 lbs (max.) |
Ship Weight: | 78 kg (max.) | 170 lbs (max.) |
Noise Output: | < 78dB |
Power Supply Requirements | |
---|---|
Voltage | 100 – 240 V automatic conversion |
Frequency | 50/60 Hz |
Power Consumption | < 300 VA |
System Specifications | |
---|---|
Number of Plates | 2 |
Sample Capacity | 100 per load |
Continuous Sample Loading | Yes |
Sample Tube Size | 10 mm – 16 mm diameter / 40 mm – 100 mm height |
Number of Reagents | Up to 8 (25 mL tubes) / Up to 10 (15 mL tubes) |
Number of Standard/Control Bottles | 24 (2 mL vials) |
Assays per Plate | Up to 12 |
Dilution Wells Capacity | Up to 96 deep-well strips (12 x 8) |
Sample Tip Capacity | 216 tips (2 x 108) |
Reagent Tip Capacity | 20 tips |
Reader Specifications | |
---|---|
Photometric Range | 0.100 – 3.0 OD |
Spectral Range | 405 nm – 690 nm |
Filter Slots | 6 wavelength |
Precision | < 1% CV (< 2.0 OD) < < 2% CV (2.0 – 3.0 OD) |
Accuracy | ± 0.005 OD or 2.5% (0.000 – 3.000 OD) whichever is greater |
Read Time | < < 30 sec (single wavelength) |
Washer Specifications | |
---|---|
Manifold Configuration | 8-way |
Programmable Volumes | 50 μL – 1000 μL |
Wash Buffer Capacity | 2 wash bottles at 2 L, with level-sensing |
Residual Wash Volume | < < 3 μL with super sweep |
Dispense Precision | ≤ 5% CV (with 300 μL in a 96 well plate) |
Waste Container | 1 x 1.5 L |
Incubator Specifications | |
---|---|
Number of Incubators: | 2 |
Temperature Range: | Ambient + 4° C to 40° C |
Temparature Accuracy | ± 1° C across plate @ 37° C |
Shaking | Independent linear motion 14 – 20 Hz (periodic or continuous) |
Temperature Monitoring | Yes |
Regent Pipetting Specifications | |
---|---|
Regent Tip Size: | 1,300 µL |
Reagent Pipetting Volume: | 20 μL – 1,000 μL |
Reagent Pipetting Precision: | < 3% CV at any operating volume above 20 uL (single-shot mode) |
Sample Pipetting Specifications | |
---|---|
Sample Tip Size | 300 µL |
Sample Pipetting Volume: | 10 µL – 250 µL |
Sample Pipetting Precision | ≤ 3% CV (at any operating volume above 10 µL) |
Dilusion Range | 1 part in 199 one-stage dilution <1 part in 39,601 two-stage dilution |
Process Security | |
---|---|
Liquid Level Sensing | Yes (reagents, controls and samples) |
Level Sensor System | Pressure Differential |
Clot Detection | Yes |
Dispense Anomaly Detection | Yes |
Tip Detection | Yes |
Electronic Signature Pipetting | Yes |
Alarms | Yes |
Software | |
---|---|
Controlling Software | DS-Matrix® |
Work Protocols (Assays): | Unlimited |
Data Processing | Quantitative & qualitative |
Lever-Jennings | Yes |
Westgards Rules | Yes |
Process Reporting | Event log + Error log |
Automatic Error Recovery | Yes |
Password Access Control | Yes |
Attributes and specifications subject to change without notice.